The "Ht" and thermometer
icon appears. Temperature is
too high, outside the required
range of 50°F ~104°F (10°C
~40°C). This alerts users that
an incorrect result may occur
if the test continues.
Action: Relocate the meter to a
location with temperature
between 50°F ~104°F (10°C
~40°C).
The "Lt" and thermometer
icon appears. Temperature
is too low, outside the
required range of 50°F
~104°F (10°C ~40°C). This
alerts users that an
incorrect result may occur
if the test continues.
Action: Relocate the meter to
a location with temperature
between 50°F ~104°F (10°C
~40°C
)
.
Test result is
higher than 600
mg/dL.
Action: Test again.
If the result is still
the same, please
contact your
healthcare
professional.
The test result is
lower than 20
mg/dL.
Action: Test again.
If the result is still
the same, please
contact your
healthcare
professional.